hm 460 AS-V / hm 660 AS-V

Validatable bar sealers (Impulse)

hm 460 AS-V / hm 660 AS-V

You have questions about our products? We are always happy to help.

Give us a call or send us an email:
T +49 (0) 6261 9770-0
sales@hawo.com

hm 460 AS-V / hm 660 AS-V

The microprocessor-controlled hm 460 AS-V / hm 660 AS-V impulse sealers are used in the packaging of sealable pouches and reels (SBS) in hospitals and the medical industry. The sealing process is validatable in accordance with ISO 11607-2 and meets the requirements of the therefrom resulting specifications of the Validation Guideline Guideline for the Validation of the Sealing Process.

The clear keyboard of the hm 460 AS-V and hm 660 AS-V devices makes menu navigation and data input exceptionally simple. Information is clearly and precisely presented on the well laid out, 2-line LCD display.

The critical process parameters of temperature, contact pressure and sealing time are monitored directly by the devices. For long-term documentation the devices are compatible with additional systems (e.g. ht 180 PT-USB).

Features

  • RS 232 interface for PC connection
  • Microprocessor-controlled sealing time automation
  • Menu-driven, multilingual input of the sealing parameters
  • Adjustable and absolute batch counter
  • Integrated film roll holder with cutting device

Safe packaging

The correct reprocessing process of medical devices consists of the steps of washing & disinfection, packaging and sterilisation. The instruments can only be called sterilised when they are packaged before the sterilisation. The single-use (!) packaging, made from laminated poly film and a porous material (Tyvek® or medical grade paper) is permeable for the sterilisation medium (e.g. steam, plasma, Formaldehyde FO or Ethylenoxide ETO), but not for bacteria or microorganisms. Only by following this reprocessing sequence and by using professional heat sealers to seal the instruments as well as professional packaging material can the sterility up to the point of use as well as the aseptic presentation of the instrument be guaranteed.

Medical devices delivered in a sterile state should be packed to ensure that they remain sterile until the point of use. The validation of packaging processes is crucial to ensure that sterile barrier system integrity is attained and will remain so until opened by the users.

The international packaging standard ISO 11607-2 as well as the Technical Specification draft ISO/DTS 16775 explain how packaging processes should be validated. The international packaging validation guideline gives guidance how to validate packaging processes. During the validation of the heat sealing process the optimum sealing temperature of the packaging materials used has to be evaluated and established. At this temperature, the process then has to deliver optimum seal seams that are strong enough and peelable.

Professional heat sealing equipment for closing sealable pouches and reels (preformed sterile barrier systems) are essential for this. hawo therefore offers sealing devices whose processes can be validated according to ISO 11607-2, the Technical Specification draft ISO/DTS 167751 and in harmony with the international packaging validation guideline. The models marked with ‘V’ feature an integrated function for regulating and monitoring process parameters – the ‘V’ is the key! Sealing devices from hawo therefore guarantee efficient and reproducible packing, even for large volumes of instruments.

Definition:

  • IQ Installation Qualification: essential information is checked according to a checklist (e.g. the sealing device is compliant to ISO 11607-2)
  • OQ Operational Qualification: the optimal sealing temperature has to be established
  • PQ Performance Qualification: determination of the seal strength at the optimal sealing temperature determined in the OQ (after sterilisation)
Download product brochure
hm 460 AS-V / hm 660 AS-V

Accessories

ht Seal Check HDPE
ht Seal Check HDPE
Keyfeatures
ht Seal Check med
ht Seal Check med
Keyfeatures
hawo InkTest
hawo InkTest
Keyfeatures
  • Quick, safe and easy to use for routine monitoring
  • Valid results within 20 seconds
ht 180 PT-USB
ht 180 PT-USB
Keyfeatures
  • External process documentation
  • PC backup via USB stick
  • Compatible with all validatable hawo sealing devices ("V")